Journal of Modern Laboratory Medicine ; (4): 152-156, 2017.
Article in Chinese | WPRIM | ID: wpr-611027

ABSTRACT

Objective To study application of the quality management of Blood Cell Analyzer according to the requirement of biological variation determination.Methods Collected the indoor imprecision value(CV%) from 8 items detected by blood cell analyzer during from April to Nov.of 2016,and the bias (Bias%) of 8 items of two EQA (external quality assessment) from the ministry of health in 2016.Then according to the 3 levels of the minium,appropriate and optimal quality specifications derived from the biological variability the rates of imprecision and bias were culculated.The pass rate of the imprecision and bias was calculated.By using mean bias and mean imprecision and biological variation 3 levels of total error (TEa) crite rion,and to calculate the corresponding σ and QGI value,so as to evaluate the performance of whole blood cell analyzer.Then improved the quality.Results For the imprecision value of 8 items,except the MCHC average value,all others were all 100 % meeting the appropriate level of quality requirements.For the bias value (Bias %) from 8 items,except MCH,all others were over 80 % meeting the appropriate level of quality requirements.While for the calculated σ value,based on the best level of quality requirements,except the σ value of WBC was 4.6,the σ value of all other items were all<3.Based on the appropriate level of quality requirements,except the σvalue of MCHC was 1.9,the value of σ of all other items were all> 3,and based on the minimal requirements,the σ value of all 8 items were all >3.After analysis,this blood cell analyzer,except that MCHC should use the minimal quality standard requirements,all other examination items could used the proper quality standard requirements,and the calculated QGI were all <0.8.Conclusion Based on the biological variation determination requirement and calculated σ and QGI value,this method could be used to more accurate quality evaluation of blood cell ana lyzer.Which is a higher levelof quality management,will be more conducive to quality improvement and better serve the clinical.

2.
Journal of Modern Laboratory Medicine ; (4): 159-161, 2014.
Article in Chinese | WPRIM | ID: wpr-476034

ABSTRACT

Objective Using Hitachi 7170 external quality assessment return target value to evaluate the accuracy of 10 items after Regression calibration of the Vistros 350 dry-type Biochemical Analyzer.Methods The same quality control samples were separately tested on two instruments,and results were reported to the clinical National Center for Clinical Laborato-ries.Substituted the external quality assessment return target value result from the National Center for Clinical Laboratories by using Vitros350 into regression calibration equation,then the getting data were compared with the external quality assess-ment return target value obtained from Hitachi 7170,and the deviation analysis was processed.The total error range from the America Clinical Laboratory Amended Bill was used as the standard.For the results within the reference range,error less than 1/2CLIA’88 total error,taken as the comparable judging standard,as it satisfied the requirement.For the results out off the reference range,error less than CLIA’88 total error,those still satisfied the requirement.For those items not meet the requirements,it must to do the regression calibration for Vitros350,using Hitachi 7170 as the standard instrument.Results The deviations of 7 items were all less than 1/2CLIA’88 allowed total error,with LDH was 0.16~-9.89,CK 2.92~6.25, ALT -4.64~-8.07,TBIL 0.08~2.67,TP -0.37~4.41,ALB 2.74~4.77 and URIC 1.04~3.0 respectively,and did not need re-calibration.For GLU and CREA,only one out of the reference range sample,the error range was >1/2CLIA’ 88,but

SELECTION OF CITATIONS
SEARCH DETAIL